##Modules
- Dead time
- Does solid angles, detector efficiency, and dead time correction.
- Parameter(s): Deadtime Constant (float)
- Generate Transmission
- Does monitor normalization and generates transmission of sample and transmission of empty beam
- Parameter(s): Monitor Count value (float), Coordinates of bottom-left and top-right (subject to change after widgets)
- Initial Correction
- COR = (SAM - BGD) - (Tsam/Temp)(EMP - BGD)
- Parameter(s): None
- Detector Sensitivity
- CAL = COR/DIV
- Parameter(s): None
- Absolute Scaling
- Converts to absolute scale
- Parameters(s): Instrument Name (string), meaning NG3,NG5, or NG7. Also coordinates.
- Annular Average
- Converts to 1D (Q vs. I)
- Parameter(s): None